Efficient AC/DC power supplies suit
medical use
The AcuQor family operates over a universal input range of 85-264V RMS and 47-63Hz, making it suitable for worldwide deployment.
SynQor has released its first AC/DC product designed specifically for medical applications The ACuQor range of power supplies packs 400W of useable power into 76 x 127 x 36.7mm
This article was originally published on Electronicstalk on 8 May 2001 at 8.00am (UK)
Related stories
60W dual-output convertor needs no heatsink
DualQor dual-output industry-standard quarter-brick-sized isolated DC/DC convertors from SynQor are capable of delivering up to 60W at 5 and 3.3V, with no need for an attached heatsink.
Convertor produces 40A from a quarter brick
SynQor has introduced a 40A quarter-brick-sized isolated DC/DC convertor.
Single outputs of 48, 36 or 24V are available to facilitate the use of distributed power architectures and battery backup and to directly drive components such as fans, motors, pumps, lamps and actuators.
The AcuQor family operates over a universal input range of 85-264V RMS and 47-63Hz, making it suitable for worldwide deployment.
Full output power is available up to 50C with natural convection (derates linearly to 300W at 70C).
A transient power rating of 500W is possible for up to 5s in the open frame version and even longer in an encased version.
Active power factor correction is incorporated to a level of over 0.98, enabling compliance with IEC/EN61000-3-2.
Both AC lines are internally fused.
The ACuQor models include versions designed for both BF (direct patient contact) and CF (direct cardiac contact) applications in accordance with UL/EN60601-1.
Input earth leakage current and type CF patient leakage current are well below the requirements of the standard at less than 110uA and less than 4uA, respectively.
There is also the option of ordering defibrillation-rated models that incorporate the minimum output to protective earth creepage and clearance distances and withstand the defibrillator pulse test guidelines of UL/EN60601-1.
Conducted noise emissions are below Level B limits of EN55011/55022, FCC part15.
A host of additional control and signalling inputs and outputs are offered, including a 5V "always on" standby power output, a DC power good signal, AC power good signal, remote enable input and a fan status output/fan enable input function.
Overcurrent, overvoltage and overtemperature protection with auto ecovery is also built into the unit.